Features Cooperate with an 0.00385 alpha
coefficient, meet DIN IEC 751.
0 ℃ resistance 100 ohm.
Measuring -50 to 400 ℃.
Range -58 to 752 ℉.
Plug DIN plug, 4 pins/4 wires.
Class A.
Accuracy ± (0.15 + (0.002 x T))℃.
T : measuring temperature.
For example :
Accuracy is 0.15 ℃ for 0 ℃ reading.
Accuracy is 0.35 ℃ for 100 ℃ reading.
Accuracy is 0.95 ℃ for 400 ℃ reading.
Dimension Sensing head - 152 mm tube,
3.2 mm Dia.
Probe length - 245 mm.
Remark TP-100 is the optional probe for TM-907A,
TM-917.