HIGH VOLTAGE PROBE LUTRON HV-40

Write a review
2,560,000 تومان
3788
Out of stock

SPECIFICATIONS

Attenuate Ratio 1 : 1000.

Input Impedance Approx. 1000 M ohm.

Output Around 1.1 M ohm. The input impedance of external voltmeter

Impedance should be 10 mega ohm.

Safety Meet CAT II 40000 V

Max. Working DCV DC 40 KV

Voltage ACV Peak AC 40 KV or 28 KV rms

( depend which values is larger ).

Accuracy DCV 1 KV to 20 KV - ± 1 %.

20 KV to 40 KV - ± 1.5 %.

ACV 1 to 28 KV rms, 50/60 Hz - ± 5 %.

Temp Coefficient Less than 200 ppm/蚓.

Operating Temperature 0 to 50 ℃ ( 32 to 122 ℉ ).

Operating Humidity Less than 80% RH.

Cable Length 1 meter.

OPERATION

Connect the plugs to the volts ( Hi ) & com ( Lo ) input terminals of your voltmeter ( or

Multimeter ). Select the desired range of voltmeter ( Attention : Do not use auto ranging ).

Whenever possible, turn the high voltage source off before making any connections. Connect

the HV probe common lead ( alligator clip ) to a good earth ground or reliable chassis ground.

SAFETY PRECAUTION & WARNING !!!

* This high voltage probe must be used by the person who are trained only. Do not work alone

when working with high voltage circuits & environment.

* For your own safety, inspect the probes for cracks & frayed or broken leads before each use.

If any defects are noted, do not use the probes.

* Hands, shoes, floor & work bench must be dry. Avoid making measurements under humid,

damp or other environmental conditions that might affect the safety of measurement

situation.

* The ground connection must always be made before the probe tip comes into contact with

the high voltage & must not be removed until after the probe tip has been removed from

high voltage source.

* Do not attempt to take measurement from sources where the chassis or return lead is not

ground.

* If possible, always turn the high voltage source off before connecting or disconnecting the

probe.

* Before turning the high voltage on, make sure that no part of your body is in contact with

the device under test.

* The probe body should be kept clean & free of any conductive contamination. Clean only

the exterior probe body & cables. Use a soft cotton cloth lightly moistened with a mild

solution of detergent & water. Do not allow any portion of the probe to be submerged at

any time.

 

 